GAMMA COLONNE III

Extremely robust, quiet running. With high-quality linear guide. Optionally equipped with potentiometer. Designed for use in hospital and care beds.

  • Dynamic load capacity: max. 2000 N as central load; max. 350 N at 400 mm lever arm length
  • Static load: max. 1000 N at 400 mm lever arm length
  • Adjustment speed: max. 14.5 mm/sec. *
  • Stroke length: 400 mm
  • 3-part aluminium column, 2-stage telescopic upwards, inner profiles telescope outwards

* depending on drive configuration at constant 24/29 V DC without load

Category:

STANDARD CONFIGURATIONS

Performance data
Dynamic load capacity max. 2000 N as central load
max. 350 N at 400 mm lever arm length
Static load max. 1000 N at 400 mm lever arm length
Adjustment speed 1 max. 14.5 mm/sec.
Stroke length 400 mm
Fitting dimension 370 mm
Duty cycle 2/18 min. or 10 %, max. 5 switching cycles per minute
Motor voltage maintenance-free 24 V DC motor
Components
Limit switch with micro switch
Guidance profile aluminium anodised
Housing colour light grey RAL 7035
Motor connecting line grey, straight, 2000 mm, LSP-plug, PVC
grey, straight, 1600 mm, 8-pole DIN-plug, PVC
grey, spiralled, 560 mm, LSP-plug, PVC
Further data
Protection type max. IPX6
Protection class III
Relative humidity 30% – 75%
Ambient temperature +10°C – +40°C

EXTRA CONFIGURATIONS

Components
Position feedback potentiometer
Further data
Assembly options base plate for direct screw connection, fitting sleeve
Equipment options anti-twist device up to max. 100 Nm

1 depending on drive configuration at constant 24/29 V DC without load