MC 15: VERSATILE AND SAFE CONTROL UNIT FOR MEDICAL APPLICATIONS

You’re on the safe side with the MC 15: This control unit features single-fault protection. It can connect up to four drives and up to two control keypads. In addition to its integrated SMPS, the MC 15 features an excellent price-performance ratio. It can also be directly mounted to the MEGAMAT 10 or MEGAMAT 5 single drive. This control unit also has IPX6 protection so that it is protected against the ingress of very small contaminants and foreign particles.

CUSTOMIZED TO YOUR NEEDS

In addition, there are many expansion options for the control unit so that it can meet requirements for hospital and nursing beds. These include a 29-volt DC connector for peripherals (such as external light sources). An optional rechargeable battery is also available so that a nursing bed can be operated in the event of a power failure. A power cable with earthing/ground is also available. An optional configuration features IPX6 protection, which ensures that the housing is protected against the ingress of solids and liquids as well as against touch-contact voltages.

Kategória:

CUSTOMER BENEFITS

rechargeable battery connection, power outlet (29 V DC) for e.g. external light systems, mains cable with earthing connection
connections for drives: 1-4
connections for operating elements: 1-2
First error safe in accordance with medical requirements and corresponds to the Eco-design guideline (power consumption <0.5 W in standby)

STANDARD CONFIGURATIONS

Standard configurations
Type Single-fault protected control unit
Protection class II
Protection type IPX6
Input voltage wide input voltage range 100-240 V AC
Transformer Switching mode power supply (SMPS)
Duty cycle 2/18 min. or 10%.max.5 switching cycles per minute
Connectivity connections for dives: MC10:3,MC11/MC15:4
Protection degree connections for operating elements:M C10:1-2, MC11/MC15:1-2
Ambient temperature +10℃-40℃
color light grey
supply cable pluggable main cable, USA, ENG, EU, CHN, AUS…
Assembly options assembled with megamat10, or with mounting plate
Certification IEC60601-1, IEC60601-2-52